Our Accelerometer Calibration Laboratory was established in 2021 and is located in TOSB BİAS Automotive Test Center. It serves due to TS EN ISO/IEC 17025:2017 General Requirements for Experiment and Calibration Laboratory quality certificate.
Our automated accelerometer calibration system includes options for calibration frequencies from 5 Hz to 10 kHz with very low uncertainty. The system allows for sensitivity and phase calibrations within this range in compliance with ISO 16063-21: 2016. The ISO 16063-21 standard, “Vibration calibration by comparison to a reference transducer” describes the calibration of rectilinear vibration transducers by comparison to a standard that is calibrated by primary methods. MB Dynamics Win475 automated accelerometer calibration system are used for accelerometer calibration processes.